Introduction Non-Invasive Blood Pressure (NIBP)
14 - 2 V Series Operating Instructions
14.1 Introduction
This chapter describes the Non-Invasive Blood Pressure (NIBP) function in detail including how to
configure NIBP alarm limits, monitor NIBP, and configure the NIBP parameter.
The V 12/V 21 calculates NIBP values using the oscillometric method of noninvasive blood pressure
measurement. The NIBP measurement includes Systolic (Sys), Diastolic (Dia) and Mean Blood
Pressure. These measurements are determined accurate after comparison to auscultatory
measurements per ANSI/AAMI SP10 standard with the fifth Korotkoff sound which is used to
determine diastolic pressure.
The V 12/V 21 measures blood pressures that are equivalent to those obtained by a trained observer
using the cuff/stethoscope auscultation method, within the limits prescribed by the American
National Standard, Electronic or automated sphygmomanometers.
Patient Sizes
Intended patient types: Adult, Pediatric, Neonate.
VPS Connector
The VPS module supports one (1) NIBP connector (as shown in FIGURE 14-1). Refer to "Measuring
NIBP" on page 14-5 for additional information.
FIGURE 14-1 Example Nellcor VPS NIBP Connector
NIBP Connector

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals