EasyManuals Logo

Mindray V21 User Manual

Mindray V21
587 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #429 background imageLoading...
Page #429 background image
System Dialog System Setup
26 - 6 V Series Operating Instructions
FIGURE 26-3 Example System Dialog (Alarms Tab)
4. Select the buttons described in TABLE 26-3 to configure the tab.
TABLE 26-3
BUTTON DESCRIPTION / COMMENTS CHOICES*
Audio Pause Timeout Allows an authorized user to choose which
settings are available in the Alarm Setup
dialog Audio Pause functions. This setting
controls for how long an alarm is muted.
1, 2, 5,10 min, Permanent
1, 2, 5, 10 min
1, 2, 5 min
1,2 min
NOTE: The Audio Pause configuration applies to all
patient sizes.
Lethal Alarms Allows an authorized user to determine
whether or not a user can modify the lethal
arrhythmia settings and decide whether
Asystole, VTach and VFib alarms can be
turned off by the user.
Locked
Unlocked
Apnea Alarms Allows an authorized user to enable Apnea
alarms at the monitor.
Locked
Unlocked
Latch Alarms Configures whether or not to latch the
physiological alarms.
All On
All Off
Lethal Only
Alarm/Response
Order
Configures the order of the Alarm Limit
dialog and Alarm Responses dialog tabs.
Refer to "Configuring the Alarm/
Response Tab Order" on page 26-
7.
Mixed Agent Audio Allows an authorized user to enable or
disable the Mixed Agent Audio.
On
Off
* The factory default configuration settings are in bold.

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals