V Series Operating Instructions 29 - 5
Product Specifications Specifications
29.1.3 V 21 Safety Designations:
TABLE 29-3
Type of protection against electric shock: Class II
Supply Connection: 14.3Amps (max), +15Vdc Input
3.7 VDC Internal Backup Battery
Mode of Operation: Continuous
Protection Against Hazards of Explosion: Not protected (Ordinary)
Protection Against Ingress of Liquids Drip-proof (IPX1)
Degree of electrical connection between
equipment and patient
Equipment designed for direct electrical and non-electrical
connection to the patient
Degree of Mobility: Stationary
Maximum Size: 23"H x 16"W x 9"D
Maximum Weight: 23 lbs.
Operating Temperature: +5°C to +40°C
Operating Humidity: 15% to 95%, non-condensing
Operating Altitude: 1060hPa to 700hPa
Storage Temperature: -20°C to +60°C
Storage Humidity: 10% to 95%, non-condensing
Storage Altitude: 1060hPa to 570hPa
Controls/Indicators: 1- Touchscreen
1- 21.3" TFT Display
2- Alarm LED
1 - Touchpad
1- AC Present
1- Battery(ies) Charging
Display: Operates in portrait orientation
Minimum +50/-50 degree horizontal viewing angle
Minimum +45/-45 degree vertical viewing angle
Aspect ratio of 3:4 in portrait mode
Display resolution is 1600 x 1200 pixels
Interfaces: 1- Diagnostic Port (DP1)
1- V Dock Interface (D1)
1- V Hub Interface (MR4)

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals