4-7 
 
4.5.2 N2O Cylinder Supply Test 
1.  Remove the N2O supply hose from the gas inlet assembly. 
2.  Mount the backup cylinder full of N2O on the backup cylinder yoke. If necessary, place 
a new clean washer between the cylinder and cylinder yoke to minimize leakage caused 
by yoke connections. 
3.  Keep the N2O cylinder open, before the cylinder pressure gauge displays the pressure of 
the cylinder. 
4.  Turn off N2O cylinder. 
5.  Ensure that the pressure value displayed on the front panel N2O pressure gauge does not 
decrease by less than 10% within one minute. 
 
4.5.3 Air Cylinder Supply Test 
1.  Remove the Air supply hose from the gas inlet assembly. 
2.  Mount the backup cylinder full of Air on the backup cylinder yoke. If necessary, place a 
new clean washer between the cylinder and cylinder yoke to minimize leakage caused 
by yoke connections. 
3.  Keep the Air cylinder open, before the cylinder pressure gauge displays the pressure of 
the cylinder. 
4.  Turn off Air cylinder. 
5.  Ensure that the pressure value displayed on the front panel Air pressure gauge does not 
decrease by less than 10% within one minute. 
 
4.6 EFCS Accuracy Test 
Switch the system to the ACGO mode and connect the pipeline from the separate ACGO to the 
test equipment (Alicat is recommended). 
Verify the flow accuracy based on the following table. 
Test Item  Testing Method 
Acceptance 
Criterion (Flow 
Value Displayed 
on Alicat) 
(Direct flow control mode) 
Accuracy of the O2 
electronic flowmeter 
(Balance gas setting: 
None) 
Gas is set to O2 on 
Alicat 
Adjust the O2 flow knob to set the flow to 
0.2 L/min. 
0.2 ± 0.05 L/min 
Adjust the O2 flow knob to set the flow to 
1 L/min. 
1.0±0.05 L/min 
Adjust the O2 flow knob to set the flow to 
3 L/min. 
3.05±0.15 L/min 
Adjust the O2 flow knob to set the flow to 
15 L/min. 
15.25±0.75 L/min 
Accuracy of the air 
electronic flowmeter 
(Balance gas setting: 
Air) 
Gas is set to air on 
Alicat 
Adjust the air flow knob to set the flow to 
15 L/min. 
15.25±0.75 L/min 
Adjust the air flow knob to set the flow to 
3 L/min. 
3.05±0.15 L/min 
Adjust the air flow knob to set the flow to 
1 L/min. 
1.0±0.05 L/min 
Adjust the air flow knob to set the flow to 
0.2 L/min. 
0.2 ± 0.05 L/min