9.2.19 Remove Flowmeters
1. Remove the EFCS flowmeter assembly.
2. Unscrew the eight (8) screws fixing the flowmeters.
3. Remove the connected cables to take out the two flowmeters.
4. Unscrew the six (6) screws fixing the outlet block assembly to take out the inlet and
outlet block assemblies.
5. Unscrew the four (4) screws fixing the flowmeter to take out the flowmeter on the outlet
block.