8.8.2 Software Function Activation
The manufacturer provides activation codes for activating the functions listed in the table below. If
the customer needs to add any of the following functions, the after-sales service engineer can
apply to the manufacturer for the activation code.
Configuration Item Function Description
PCV Pressure control ventilation mode
PCV-VG Pressure control ventilation-volume guarantee
SIMV-VC Synchronized intermittent mandatory ventilation-volume control
SIMV-PC Synchronized intermittent mandatory ventilation-pressure control
SIMV-VG Synchronized intermittent mandatory ventilation-volume
guarantee
PS Pressure support ventilation mode
CPAS/PS Pressure support ventilation/continuous positive airway pressure
ventilation
Bypass CPB mode
Spirometry Loop
8.8.2.1 Applying for Activation Codes
Note
To apply for activation codes, provide relevant information about the anaesthesia
machine for which functions are to be activated, including the machine ID, existing
configuration, and configuration to be activated.
When a customer needs to add pad configurations, the after-sales service engineer must apply to
the after-sales service department of Mindray for the corresponding activation codes. The
procedure is as follows:
1. Record the serial number (see the label on the left of the machine) of the anaesthesia
machine for which configurations are to be activated.
2. Record the machine ID and existing configurations. Choose Main > System >
Configuration Information to access the Configuration Information menu.