7-52
7.3.7.3 Calibration (Factory) Procedure
Calibrate the flow sensor as follows:
1. Ensure that the machine is in standby mode.
2. Select [Main] > [Service] > [Calibrate] > [Calibrate Total Flow Sensor] to go to the
screen as shown in the following figure:
3. Select [Begin]. A screen as shown in the following figure is displayed.