9-1
9 Repair and Disassembly
WARNING
Only use lubricants approved for anesthesia or O2 equipment.
Do not use lubricants that contain oil or grease. They can burn or explode in the
presence of high O2 concentrations.
Obey infection control and safety procedures. Used equipment may contain blood
and body fluids.
Movable parts and removable components may present a pinch or a crush hazard.
Use care when moving or replacing system parts and components.
Be careful with edges or corners of mechanical components when disassembly, to
avoid cuting.
Be careful with screws, to avoid screws falling into the machine to cause short
circuit.
Before remove gas devive, ensure to release the pressure of gas to avoid personal
injury caused by high pressure.
NOTE
Check whether spare parts are demaged, and replace them if necessory, during the
process of reinstallation. Choose and use screws and spare parts correctly.
After repair or replacing, check and test according to the chapter 4 Testing.