8-41 
Failure Description  Possible Cause  Recommended Action 
gauge is damaged.    pressure gauge. 
When the O2 supply 
pressure is too low, 
the "O2 Supply 
Failure" alarm is not 
generated; or when 
the O2 supply 
pressure is within 
the specified range, 
the "O2 Supply 
Failure" alarm is 
generated.  
The pressure switch of 
the O2 inlet assembly is 
faulty.  
Adjust the pressure switch of the O2 inlet 
assembly to keep the O2 supply pressure 
between 0.15 MPa and 0.25 MPa, as close to 
0.2 MPa as possible when this alarm is 
generated. If the adjustment fails, replace the 
pressure switch (refer to section 
8.6.1.3Adjusting the Pressure Switch).  
When the drive gas 
supply pressure is 
too low, the "Drive 
Gas Pressure Low" 
alarm is not 
generated; or when 
the drive gas supply 
pressure is within 
the specified range, 
the "Drive Gas 
Pressure Low" alarm 
is generated.   
The pressure switch of 
the integrated airway of 
the exhalation valve or 
PEEP safety valve is 
faulty. Alternatively, the 
filter of the integrated 
airway of the exhalation 
valve assembly is 
blocked.  
Adjust the pressure switch of the integrated 
airway of the exhalation valve assembly to 
keep the drive gas pressure between 0.05 
MPa and 0.2 MPa, as close to 0.14 MPa as 
possible when this alarm is generated. If the 
adjustment fails, replace the pressure switch. 
If the issue persists after the pressure switch 
is replaced, replace the integrated airway of 
the exhalation valve assembly (refer to 
section 8.6.1.3Adjusting the Pressure 
Switch).  
 
8.4.2.1  Testing the Pipeline Pressure Gauge and Calibrating the 
Regulator 
Use the following tools to test the pipeline pressure gauge and regulator of the pipeline inlet 
assembly: 
  One 1 MPa (10 bar) test pressure gauge (before test, ensure that the test pressure gauge 
is in good conditions) 
  One 3106-04-06 adapter connector  
  One PU tube (4X200) 
  One PU tube (6X200)