7-2
7.2.2 Cautions
CAUTION
Refer to the maintenance period in the Chapter 6 Device Maintenance when
conducting periodic maintenance.
If the pipeline gas supply is in use and the main switch of the system is set to the
ON position, ensure that the standby gas cylinder valve is not opened. Otherwise,
the cylinder gas supply could be depleted, resulting in an insufficient reserve in the
case of a pipeline gas supply failure.
Use the cleaning agent sparingly. Excess fluid could enter the machine, causing
damage.
The machine can be operated only by trained and skilled medical personnel.
7.2.3 Notes
NOTE
Only bacterial filter with a low-flow resistance can be connected to the patient
module and/or patient pipe.
Wear surgical gloves when touching or disassembling valves or other internal
components of the breathing system.
Ensure that the gas supply of the machine always complies with technical
specifications.
The APL valve and the PAW gauge marker are for reference only. The anesthesia
machine displays the calibrated airway pressure.
If the machine malfunctions during initial calibration or testing, do not use it until
the fault is rectified by a professional service technician.
After servicing, functional tests, sensor tests, and system tests must be carried out
prior to clinical application.
The machine supports up to two vertical mounting brackets to accommodate
additional monitors and other devices. Unauthorized mounting accessories are not
recommended.
Ensure that all devices on the top plate of the machine are secured.