3-1
3 Installation Guide
3.1 Preparation
Before installation, contact the consumer and tell him/her to provide the following materials. The
customer is responsible for supplying these materials. Missing items may result in delays,
incomplete installations, and/or additional service visits.
Gas supply hose connector.
O2 cylinder, N2O cylinder, and Air cylinder in compliance with the configuration
standard.
Anesthetic vaporizer and adapter (if not purchased together with the anesthesia system)
Anesthetic agent
Loose fill or Pre-pak CO2 absorbent
Effective O2, N2O, and Air gas sources (280 to 600kPa (40 to 87psi))
Dropdown hoses for ceiling mounted medical gas equipment, compatible with
quick-disconnect hoses ( if not purchased together with the anesthesia system)
3.2 Assembly
NOTE
After the flow sensors are calibrated, the anesthesia machine must be used together
with the breathing system module. If the breathing system module is removed,
make sure that it is re-installed onto the original anesthesia machine. If other
breathing system module is to be installed, the flow sensor must be re-calibrated
(please refer to 7.3.2Flow calibration (Service)).