7-47
5. The interface as shown in the following figure is displayed after the calibration is
complete. Select [OK] to quit calibration.
7.3.6.2 100% O2 Calibration
NOTE
100% O2 calibration must be performed in standby mode.
100% O2 calibration can be performed only after 21% O2 calibration is completed
successfully.
Perform 100% O2 sensor calibration as follows:
1. Enter the standby mode.
2. Select [Main] > [System] > [Calibrate] > [O2 Sensor] or [Main] > [Service] >
[Calibration] > [O2 Sensor]. The [System] and [Maintenance] tab pages require the
password for 21% and 100% O2 sensor calibration. Select [100%]. The calibration
interface as shown in the following figure is displayed. Set the machine according to the
instructions displayed on the screen, and select [Next]. Set the Auto/Manual switch to
the Auto position.