9-28
9.2.28 Remove Gas Reservoir Assembly
1. Remove the cover plate of work surface.
2. Disconnect the pipeline and cables connected with the gas reservoir.
3. Unscrew the two (2) screws to remove the gas reservoir.
9.2.29 Remove O2 Flush Assembly
1. Remove the cover plate of work surface.
2. Unscrew the four (3) M3X8 combined screws and the four (4) M3X8 countersunk head
screws fixing the front panel bracket, to remove the front panel bracket.
3. Unscrew the four (4) M3X8 combined screws fixing the O2 flush bracket and the four (4)
screws fixing the O2 flush assembly on the bracket, to remove the O2 flush assembly.