8-40 
 
Ensure no leakage occurs before using the negative pressure ball. Check whether the pre-seal 
cover is fastened. Then, squeeze the negative pressure ball to exhaust all gas inside. Properly 
install the exhaust port plug. Block the pre-inlet port using your fingers and then release the 
negative pressure ball. Ensure that the negative pressure ball is obviously expanded within 30s. If 
not, replace the negative pressure ball.   
 
8.4.2  Effectiveness Test for the Pipeline Pressure Gauge and 
Correction of the Regulator 
The following table lists the faults related to gas source and drive gas.   
Failure Description  Possible Cause  Recommended Action 
Leakage 
The supply pipeline is 
damaged or the seal ring 
of the connector is 
damaged.  
Replace the supply pipeline or seal ring of the 
connector. 
The quick connector is 
leaked.  
Replace the quick connector or PU tube. (If 
the PU tube is not damaged and has enough 
length, cut a certain segment of the PU tube 
that is in contact with the quick connector and 
insert the segment into the PU tube.) 
The pipeline inlet 
assembly is leaked.   
Check whether the check valve of the 
pipeline gas source assembly has a leakage 
issue in the reverse direction. If necessary, 
replace the check valve. Check whether the 
seal ring on the pipeline inlet assembly is 
damaged. If yes, replace the seal ring. If the 
issue persists, replace the pipeline inlet 
assembly.  
The drive gas pipeline is 
leaked.  
Check and repair the exhalation valve 
according to section 5.3.4.2 Low Pressure 
Airway System Leakage Test. 
The pipeline 
pressure gauge 
displays inaccurate 
reading or no 
reading. 
The pipeline pressure 
gauge is damaged.   
Replace the pipeline pressure gauge. 
The reading of the 
pipeline pressure 
gauge is sharply 
fluctuated.  
The filter of the pipeline 
supply assembly or PU 
tube of the pipeline 
pressure gauge is 
blocked, or the pressure 
If the pipeline pressure is stable, check the 
PU tube of the pipeline pressure gauge or the 
filter of the pipeline supply assembly. If the 
PU tube or filter is blocked, replace it.   
2. If the issue persists, replace the pipeline