8.8 Software Upgrade and Software Configuration
Activation
Note
After replacing the CPU board, ventilation protection board, ventilation control
board, power board, expiratory valve assembly, upgrade the software.
Before upgrading WATO EX-55/65, read the documents attached with the upgrade
package and perform the upgrade according to the documents.
1. Connect the network ports of the PC and anaesthesia machine by using a crossover
network cable.
2. Before running the Mindray anaesthesia machine software upgrade tool, check that the
IP address is set to 192.168.23.1 and the subnet mask is set to 255.255.255.0 for the PC.
Check and set the IP address of the PC. See the figure below.
3. Ensure that the Mindray anaesthesia machine software upgrade tool is installed on the
PC. If the upgrade tool is not installed, perform the following steps:
(1) Run SystemUpdateTool.exe.
(2) If a dialog box is displayed prompting you to set the language, select English and
click OK.