8-92 
Failure 
Description 
Possible Cause  Judgment Method  Recommended 
Action 
displays a 
black screen 
and the alarm 
indicator is off. 
analysis. 
The CPU board is 
abnormal. 
Power on the anaesthesia machine. If 
the sound of the valves can be heard 
and the top lamp works properly, the 
CPU board is abnormal. Otherwise, 
proceed with the analysis. 
Replace the CPU 
board. 
The power board 
is abnormal. 
If there is no response when you try to 
power on the anaesthesia machine, 
see the fault analysis at the beginning 
of this table. 
Replace the cable 
or power board. 
The 
anaesthesia 
machine 
displays a 
blank screen or 
black screen. 
The cable is not 
connected 
properly. 
Check whether the LVDS and 
backlight interface are normal. If so, 
proceed with the analysis. 
Connect the cable 
properly. 
The mother board 
is in poor 
condition. 
Open the cover plate, power on the 
anaesthesia machine, and test pin 19 
of mother board J3. If the voltage of 
this pin is beyond the range 3.0-3.6 V, 
the mother board is    poor condition. 
Otherwise, proceed with the analysis. 
Replace the 
mother board. 
The CPU board is 
in poor condition. 
If the fault cannot be identified by 
using the preceding two methods, the 
CPU board is faulty. 
Replace the CPU 
board. 
The segment 
code screen is 
not displayed 
or is displayed 
incorrectly. 
The key control 
board is in poor 
condition. 
If the segment code screen does not 
display any content and is dark, the 
key control board is in poor 
conditions. Otherwise, proceed with 
the analysis. 
Replace the key 
control board. 
The electronic 
flowmeter is in 
poor condition. 
If the segment code screen is 
displayed but the content is different 
from that of the float flowmeter, the 
electronic flowmeter board is in poor 
condition. 
Replace the 
electronic 
flowmeter board.