5-13
5.3 Valves Diagnostic Test Tools
The diagnostic function is available only in standby mode.
The diagnostic function includes valves test, inspiratory valve test, PEEP valve test, and safety
valve test.
During the test, when you click the [Done] button in the [Diagnostic Tests] menu, the system
returns to the [Service] menu, and the relevant test stops.
5.3.1 Valves Test
The valve test menu includes the safety valve setting, inspiratory valve flow setting, inspiratory
valve D/A setting, PEEP valve pressure setting, and PEEP valve D/A setting. For details, see the
following figure: