EasyManua.ls Logo

Mindray WATO EX-65Pro - Parts Replacement; Non-Circuit Part

Mindray WATO EX-65Pro
474 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-6
If the anesthesia machine is configured with Pre-pak circuit, replace periodically as per the
following list 115-034698-00 one year complete service kit.
S/N P/N Description Qty Usage
1 049-000240-00 pop-off rubber pad 1
PoP-off valve
rubber pad
2 082-001501-00 O-ring 27X1.5, silicone, A50 2
Valve cover for
upper cover
assembly
3 082-001503-00 O-ring 20X1.5, silicone, A50 2
Valve body for
median plate
assembly
4 082-0001504-00 O-ring 23.47X2.95, silicone, A50 1
Water collection
cup for lower
cover assembly
5 082-001508-00 O-ring 20.29X2.62, silicone, A50 1
Sealing ring for
the base of the
bellows
6.4.1.1 Parts Replacement
Non-circuit part
1. As required, replace the sealing ring (082-000162-00) between the vaporizer support
rack connector and the vaporizer every 12 months.
Sealing rings
(four)

Table of Contents