EasyManua.ls Logo

Mindray WATO EX-65Pro - Page 145

Mindray WATO EX-65Pro
474 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-45
3. When the system is standby, select Main-> Service-> Diagnostic Tests->Valves to
access the following menu.
4. Set Safety Valve to On.
5. Set PEEP Valve Pressure to the following values:
(5±1)cmH2O, (20±1)cmH2O, (50±1)cmH2O, (70±2)cmH2O, (90±2)cmH2O
6. Make sure that the deviation between the measured data of the PAW sensor, PEEP
pressure sensor and that of the anesthesia machine calibration device must not exceed 1
cmH2O or 2% of the measured value of the calibration device, whichever is greater.
Otherwise, perform pressure calibration again.

Table of Contents

Related product manuals