EasyManua.ls Logo

Mindray WATO EX-65Pro - Page 148

Mindray WATO EX-65Pro
474 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-48
Single Fault
condition(SFC)
____μA
5 Mains on Applied Part Leakage ____μA Max:5000μA
6
Patient
Auxiliary
Current
Normal
condition(NC)
____μA
Max:
NC: 100μA
SFC: 500μA
Single Fault
condition(SFC)
____μA
For periodically performance, all the test items included in the ELECTRICAL SAFETY
INSPECTION FORM shall be performed. The following table specifies test items to be performed
after the equipment is repaired with main unit disassembled.
When neither power supply PCBA,
transformer nor patient
electrically-connected PCBA is repaired or
replaced
Test items: 1, 2
When power supply PCBA or transformer
is repaired or replaced
Test items: 1, 2, 3
When patient electrically-connected PCBA When patient electrically-connected PCBA
is repaired or replaced is repaired or replaced

Table of Contents

Related product manuals