5-3
The VCM screen can monitor the measured values of flow sensors sand pressure sensors in real
time, the power supply voltage and opening of valves, the power supply voltage of the ventilator,
the status of three way valve etc.
The following figure shows the screen display of VCM A/D channel data:
When the actual value in the table is displayed in red, this value exceeds the reference range
provided in the table. Further testing is required for confirmation.