(2). Connect the Fluke VT Plus to the anesthesia machine. The following figure shows the
corresponding port:
Cable connection on the back of the anesthesia machine
Connectors and settings related to FPM are shown below.
Anesthesia machine end connection mode is same as Fluke VT Plus.
FPM dedicated calibration line
FPM interface and connection mode
5. Configure the calibration device.
If the VT is used as the calibration device, configure the VT in the following way:
(1) Start the anesthesia machine calibration device. On the startup interface, go to the sensor
heating interface (waiting for five minutes according to the prompt), and then go to the zeroing
RS232 connecting
cable, connecting the
anesthesia machine