7.3.3.3 Calibration Procedure
Calibrate the pressure sensor and the PEEP valve as follows:
1. Ensure that the anesthesia machine is in standby mode.
2. Select [Main] > [Service] > [Calibration] > [Pressure Sensors] to go to the screen as
shown in the following figure:
3. Select [Auto Calibrate]. The menu as shown in the following figure is displayed.