7-38 
Failure 
Description 
Possible Cause  Recommended Action 
[00008000] is 
displayed. 
The ACGO switch is in ON 
position.  
1. Check whether the ACGO switch is 
in OFF position.   
2. Check the monitoring status of the 
ACGO switch on the interface of the 
anesthesia machine.   
[00010000] is 
displayed. 
The valve resolution is 
insufficient. 
1. Check whether the gas supply 
volume is sufficient for the whole 
calibration process.   
2. Check whether the calibration 
device works properly.   
3. Replace the PEEP valve. 
[00020000] is 
displayed. 
The maximum value cannot be 
found.  
1. Use the valve diagnosis tool to 
diagnose: (1) Turn on the PEEP valve 
in 4000 DA and check that the pressure 
measured by the VT reaches 90 cm 
H2O or higher.   
(2) Turn off the PEEP valve. Open up 
the PEEP valve gradually in DA. The 
AD value measured by the flow 
pressure sensor increases gradually.   
If the above two conditions are met, a 
re-calibration is recommended.   
2. Replace the PEEP valve. 
[00040000] is 
displayed. 
The minimum value cannot be 
found.  
1. Use the valve diagnosis tool to 
diagnose: (1) Open up the PEEP valve 
from 0 to 2000 DA. If the AD value 
measured by the inspiratory pressure 
sensor increases gradually, a 
re-calibration is recommended. 
2. Replace the PEEP valve. 
[00080000] is 
displayed. 
The zero point of the calibration 
device exceeds the standard.   
1. Zero the calibration device.   
[00100000] is 
displayed. 
The small pressure calibration is 
not monotonic.   
1. Check the connection between the 
calibration device and the sampling 
line.  
2. Replace the PEEP valve assembly.