8-11
Comm Stop establish communication
with the flowmeter board
within consecutive 10
seconds.
2. Check the cable connection between
the flowmeter board and the main
board.
3. Replace the flowmeter board when
necessary.
4. Replace the main board when
necessary.
Backup Flow
Control Error
Medium 1. The voltage of the BFCS
two-way valve is out of
range.
2. When the gas supply
pressure is within the
normal range, the values in
the flow sensor in O2 limb
and in the flow sensor in
total limb are smaller than
0.5 L/min for consecutive 2
seconds in the BFCS state.
1. Restart the machine.
2. If the position sensor is faulty,
replace the BFCS assembly.
3. If the supply voltage of the BFCS
LED is out of range, check the supply
voltage and cable connection.
4. Replace the power board when
necessary.
BFCS Key Is
Stuck
Medium The BFCS hard key is in
the pressed down state for
consecutive 30 seconds.
1. Check the BFCS hard key.
2. Restart the machine.
O
2
Supply
Failure
High O
2
supply is insufficient for
the machine with electronic
gas supply pressure
monitoring.
Connect to the O
2
supply.
Air Supply
Failure
High/Medium For the machine with
electronic gas supply
pressure monitoring: A
high-priority alarm will be
triggered by low air supply
pressure if the driving gas is
air; a low-priority alarm
will be triggered by low air
supply pressure if the
driving gas is oxygen.
Connect to the air supply.
N
2
O Supply
Failure
Medium N
2
O supply is insufficient
for the machine with
electronic gas supply
pressure monitoring.
Connect to the N
2
O supply.