8-15 
Message 
Alarm 
Priority 
Cause  Solution 
should disappear. If the alarm does not 
disappear, the pressure switch fails. In this 
case, replace the pressure switch. If the 
pressure switch does not fail, check the 
connection between the pressure switch and the 
VCM board.   
3. If both of the preceding items are normal, 
replace the VCM board.   
O2 Supply 
Failure 
High 
The O2 supply 
pressure is low. 
Use the same method for solving the problem 
of low drive gas pressure to check the O2 
pressure switch.   
Drive Gas 
Switch Valve 
Failure 
High 
The control status of 
the drive gas valve is 
different from the 
detected status.   
1. On the [Diagnosis] menu, switch the drive 
gas valve to check whether the normal status is 
restored.  
2. Restart the machine for a self test again.   
Fresh Gas 
Flow Too High 
Low 
The set fresh gas flow 
is greater than 2 
L/min, and exceeds 
the gas delivery flow 
0.2 L/min. 
Reduce the fresh gas flow.   
ACGO 3-way 
Valve Failure 
Medium  The ACGO 3-way 
valve status is in error. 
1. Check the connection of the 3-way valve.   
2. Replace the 3-way valve assembly.   
3. Replace the VCM board.   
Automatic 
Ventilation 
Disabled 
Low  The power-on self test 
fails, and the result is 
"Manual Only". 
1. Restart the anaesthesia machine. 
2. If the problem persists, check the 
corresponding module based on the result of 
the power-on self test.   
3. If the problem persists, replace the 
corresponding module based on the result of 
the power-on self test.   
Auto 
Ventilation 
Disabled-Leak 
Test Failed 
Low  The automatic circuit 
leak test fails. 
1. Check whether the circuit and pipeline are 
correctly mounted. (Check whether the 
sampling port and the drain valve are blocked.) 
2. Conduct the leak test again.   
Auto 
Ventilation is 
Non-Functional 
High  The machine can only 
be manually 
ventilated, but the 
Auto/Manual switch is 
in Auto position. 
1. Set the Auto/Manual switch to the Manual 
position. 
2. Restart the machine for a self test again.   
Pressure 
Monitoring 
  For VPM: The 
monitored value of 
1. Check whether the AD of the zero point of 
each sensor is within the valid range. If the AS