8-21 
Message 
Alarm 
Priority 
Cause  Solution 
C High Imped.  electrode C is too high.  in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
C Lead Off 
Low 
BIS electrode C is 
detached from the patient.   
1. Check whether BIS electrode C is 
in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
LE High 
Imped. 
Low 
The impedance of BIS 
electrode LE is too high. 
1. Check whether BIS electrode LE 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
LE Lead Off 
Low 
BIS electrode LE is 
detached from the patient.   
1. Check whether BIS electrode LE 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
LT High 
Imped. 
Low 
The impedance of BIS 
electrode LT is too high. 
1. Check whether BIS electrode LT 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
LT Lead Off 
Low 
BIS electrode LT is 
detached from the patient.   
1. Check whether BIS electrode LT 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
RE High 
Imped. 
Low 
The impedance of BIS 
electrode RE is too high. 
1. Check whether BIS electrode RE 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
RE Lead Off 
Low 
BIS electrode RE is 
detached from the patient.   
1. Check whether BIS electrode RE 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
RT High 
Imped. 
Low 
The impedance of BIS 
electrode RT is too high. 
1. Check whether BIS electrode RT 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS Electrode 
RT Lead Off 
Low 
BIS electrode RT is 
detached from the patient.   
1. Check whether BIS electrode RT 
is in proper contact with the skin. 
2. Replace the BIS sensor. 
BIS DSC Err  Low 
BIS DSC communication 
error 
4. Replace the BIS module.   
BIS DSC Malf  Low 
The BIS DSC is faulty 
and is automatically shut 
down. 
4. Replace the BIS module.   
BIS No Cable  Low 
The BIS cable is not 
connected.  
1. Check the BIS cable. 
2. Replace the BIS cable.   
BIS No Sensor  Low 
The BIS sensor is not 
connected.  
1. Check the BIS sensor. 
2. Reinstall the BIS sensor. 
 
BIS Sensor 
Too Many 
Low 
The BIS sensor is 
excessively used. 
1. Check whether the use times of 
the BIS sensor exceeds the limit