8-27 
2.  Set the machine following the instructions on the screen, and then select Continue to 
perform the automatic circuit leak and compliance test. Progress of the automatic circuit 
leak and compliance test is displayed on the screen. As shown in the figure below, select 
Cancel to cancel the ongoing test. 
3.  After the test is completed, confirm whether the automatic circuit leak and compliance 
test is passed following the instructions on the screen. If not, deal with it following 
"Help" on the screen and the following "Note". 
NOTE 
  If the leak test is not passed, check all possible sources of leaks, including bellows, 
respiratory system tubes and CO2 absorber canister. Check whether the sources of 
leaks are properly connected and their connectors are damaged. 
 
8.3.2 Manual Circuit Leak Test 
This test checks if the pneumatic circuit has leakage in manual ventilation mode. The test items 
include APL valve, check valve, CO2 absorber canister, patient tube, flow sensor, and flow sensor 
connector. 
 
The respiratory system leak test under manual ventilation mode shall be performed following the 
steps below: 
 
1.  If the system is being powered on, the system automatically initiates a self-test and 
enters the Manual Circuit Leak and Compliance Test screen. Or, from the main screen, 
select General > Test Leak/Compliance. Select Manual Circuit Leak Test to enter the 
Manual Circuit Leak and Compliance Test screen as follows.