8-31
Name Quantity P/N
Φ6 silicone tube 3 A21-000007---*
A5 maintenance manual / 046-001140-00
Adult test lung / 0138-00-0012
Wrench / 0367-00-0080
15 mm Y-shape connection accessory / 0103-00-0508
15 mm silicon air tube (0.6 meters) 2 0004-00-0076
2.3 L silicon breathing bag / 0992-00-0139
Adjuster calibration hose / 0453-00-1216
A5 fault locating kit / 115-009450-00
Vaporizer guide / /
Dempsey430 safety analyzer or similar device / /
31/2 digital voltmeter / /
Anesthetics (N2O) analyzer with the range of ±0.3V/V%+5% / /
BCBiomedicalDPM-2301751NMC digital pressure gauge or
similar device
/ /
DISS connector with the supply press of O2, N2O, and air being
at least 35 psi
/ /
Yoke connector of the FullPISS cylinder of the O2, N2O, and air / /
Metric hexagon screw wrench kits (manual operation tools) / /
Air flow analyzer with precision of 2% / /
Communication box (required for the FlukeVT Plus gas flow
analyzer)
/ /
Lucer adapter connector / /
Network jumper cable / 0012-00-1392-06
USB flash drive / 0992-00-0297-01
* indicates that the item is a part of the 115-009450-00A5 fault locating kits