8-87
8. Turn on the pipeline air source and record the reading on the test pressure gauge. If the
difference between the readings of the test pressure gauge and pipeline air pressure
gauge exceeds 0.1 MPa (1.0 bar), the pipeline air pressure gauge is damaged. Handle
this problem according to the fault identification table.
9. Reconnect PU tube 41 to the pressure gauge.
8.6.1.2 Testing the Pressure Switch
Check that the pressure switches of the O2 supply inlet and expiratory valve comply with the
regulations by using the following tools:
1 MPa (1.0 bar) test pressure gauge (quantity: 1)
3106-04-06 adapter connector (quantity: 1)
3106-06-08 adapter connector (quantity: 1)
3140-08-00 Y-shape tube (quantity: 1)
PU tube (8x200) (quantity: 2)
PU tube (6x200) (quantity: 1)
PU tube (4x200) (quantity: 1)