9-73
3. Unscrew the three (3) screws as show in the picture.
4. Remove the O-Ring and pull out the axis pin.
5. Remove the compression spring and replace the two seals (0030-10-13077).
Take out the seal and
pull out the axis pin
Axis
Compression spring
Seal