Image Optimization 5-49
5.14.4 Cine Review
Press <Freeze> or open a elastography imaging cine file to enter cine review status.
5.15 Contrast Imaging
Contrast imaging is an option. The probes 35C50P, 3C5P and C6-2P supports the contrast
imaging mode.
Caution:
1.
Set MI index by instructions in the contrast agent accompanied
manual.
2.
Read contrast agent accompanied manual carefully before using
contrast function.
NOTE:
Make sure to finish parameter setting before injecting the agent into the patient to
avoid affecting image consistency. This is because the acting time of the agent is
limited.
The applied contrast agency should be compliant with the relevant local
regulations.
The long-term usage of contrast destruction may damage the patients tissue. If the
contrast destruction fails, and the MI surpasses the standard, it may produce the
damage to the patient. Please follow ALARA principle.
5.15.1 Basic Procedures for Contrast Imaging
To perform a successful contrast imaging, you should start with an optimized 2D image and
have the target region in mind. To perform a contrast imaging:
1. Select an appropriate probe, and perform 2D imaging to obtain the target image, and then
fix the probe.
2. Click [Contrast] on the lower right part of the screen to enter the contrast imaging mode.
3. Adjust the acoustic power experientially to obtain a good image.
Click [Dual Live] to be “On” to activate the dual live function. Observe the tissue image to
find the target view.
4. Inject the contrast agent, and set [Timer 1] at “ON” to start the contrast timing. When the
timer begins to work, the time will be displayed on the screen.
5. Observe the image, use the user-defined key to save the images. Press <Freeze> to end
the live capture.
Perform several live captures if there are more than one interested sections.
6. At the end of a contrast imaging, set [Timer 1] as “OFF” to exit the timing function.
Perform procedures 3-5 if necessary.
For every single contrast imaging procedure, use [Timer 2] for timing.
If necessary, activate destruction function by clicking [Destruct] to destruct the micro-
bubbles left by the last contrast imaging; or to observe the reinfusion effect in a
continuous agent injecting process.
7. Exit contrast imaging.
Press <B> button to return to B mode.