Image Optimization 5-27 
5.12.3.4  Measurement, Comment, and Body Mark 
In iScape image viewing status, you can perform measurement, comment, and body mark. 
The operations are the same as that of B mode. 
 
1.  The measurement accuracy for the spliced image may be degraded, exercise 
caution when measurements are performed on an iScape image.   
2.  If there is a trace during the retracing, please don’t perform measurement 
across the trace.   
5.12.3.5  Evaluate Image Quality 
Many variables may affect the overall image quality. It is important to evaluate the image 
content and quality before an image is used for diagnosis or measurements. 
 
iScape panoramic imaging is intended for well-trained ultrasound operators or 
physicians. The operator must recognize image artifacts that will produce a sub-
optimal or unreliable image. 
The following artifacts may produce a sub-optimal image. If the image quality cannot meet the 
following criteria, you shall delete the image and do image acquisition again. 
  The image must be continuous (no part of an image moves suddenly or disappears.) 
  No shadow or absent signal along the scan plane. 
  Clear profile of anatomy through the entire scan plane without distortion.     
  Skin line is continuous. 
  The images are captured from the same plane. 
  There is no large black area in the image. 
5.12.4  Cine Review 
Click [Review Cine] on the image menu in panoramic image viewing status to enter cine 
reviewing mode. In cine reviewing mode, there is a green frame marker indicating the 
sequence of currently reviewed image in the panoramic image on the left side of screen.   
In cine review status: 
  Roll the trackball to review the captured images frame by frame. 
  Click [Auto Play] to start or end auto play. 
  In auto play mode, click [Auto Play] on the image menu; or, press/rotate the 
corresponding knob to change the play speed.   
  Review to a certain image; click [Set First Frame] to set the start point. Review to 
another image; click [Set End Frame] to set the end point. In auto play mode, the 
review region is confined to the set start point and end point. 
  Click [Return] item on the image menu to exit the cine review mode, and there 
displays the panoramic image. 
  In cine review mode, press <Freeze> on the control panel to return to the acquisition 
preparation status.