Safety Precautions 1-3 
Before cleaning the system, disconnect the power cord 
from the outlet. System failure and electric shock may 
result.
 
This system is not water-proof designed. Do Not use this 
system in any place where water or any liquid leakage 
may occur. If any water is sprayed on or into the system, 
electric shock 
or device malfunction may result. If water is 
accidentally sprayed on or into the system, contact 
Mindray Customer Service Department or sales 
representative.
 
6.  Do not use a transducer that has a damaged, scratched 
surface, or exposed wiring of any kind. Immediately stop 
using the transducer and contact Mindray Customer 
Service Department or sales representative. There is risk 
of electric shock if using a damaged or scratched 
transducer.
 
DO NOT allow the patient to contact the live parts of the 
ultrasound system or other devices, e.g. signal I / O 
ports. Electric shock may occur.
 
Do not use an aftermarket probe other than those 
specified by Mindray. The probes may damage the 
system causing a profound failure, e.g. a fire in the worst 
case.
 
Do not subject the transducers to knocks or drops. Use 
of a defective transducer may cause an electric shock. 
Do not open the covers and front panel of the system. 
Short circuit or electric shock may result when the 
system hardware is exposed and powered on.
 
Do not use this system when any digital device such as a 
high-frequency electrotome, high-frequency therapeutic 
device or defibrillator is applied already. Otherwise, there 
is a risk of electric shock to the patient.
 
12.  When moving the system, you should first fold the 
keyboard, disconnect the system from other devices 
(including probes) and disconnect the system from the 
power supply.
 
Accessory equipment connected to the analog and digital 
interfaces must comply with the relevant IEC standards 
(e.g., IEC 60950 information technology equipment safety 
standard and IEC 60601-1 medical equipment 
standard).Furthermore, all configurations must comply 
with the standard IEC 60601-1. It is the responsibility of 
the person, who connects additional equipment to the 
signal input or output ports and configures a medical 
system, to verify that the system complies with the 
requirements of IEC 60601-1. If you have any questions 
regarding these requirements, consult your sales 
representative.