13-14 Probes and Biopsy
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
2P2P
0℃~40
30%85%RH
no
condensation
700hPa1060hPa
V10-4BP
0℃~40
30%85%RH
no
condensation
700hPa1060hPa
7LT4P
0℃~40
25%90%RH
no
condensation
550hPa1060hPa
6LE7P
0℃~40
30%85%RH
no
condensation
700hPa1060hPa
P7-3P
10℃~40
30%85%RH
no
condensation
700hPa1060hPa
35C50P
0-40
30%-85%RHno
condensation
700-1060hPa
75L38P
0-40
30%~85%RHno
condensation
700 hPa ~ 1060 hPa
D6-2P
18℃~30
20%85%RH
no
condensation
700hPa1060hPa
C6-2P
0℃~40
20%85%RHno
condensation
700hPa1060hPa
7L4BP
0℃~40
20%85%RHno
condensation
700hPa1060hPa
Storage and transportation Condition
When all examinations for the day have been completed, confirm that the transducer is in
good condition. After disinfecting the transducer, confirm that the transducer is in good
condition and store it in a suitable place so that the next examination can be conducted
smoothly.
Store and transport the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
3C5P
-20℃~55
30%~95%RHno
condensation
700 hPa ~ 1060 hPa
6C2P
-20℃~55
30%~95%RHno
condensation
700 hPa ~ 1060 hPa

Table of Contents