5-34 Image Optimization 
5.13.3.2  Static 3D Acquisition Preparation 
Description of parameters: 
Parameter 
adjusting 
Angle 
Function: to set the range for imaging. 
Range: 10-70°. 
Quality 
Function: to adjust the image quality by changing the line 
density. Image quality can affect the imaging speed: the 
better the image quality, the longer the time.   
Range: Low2, Low1, Mid, High1, High2 
Render Mode 
Surface 
Function: set Surface as the 3D image rendering mode. 
This is useful for surface imaging, such as fetus face, hand 
or foot. 
Tip: you may have to adjust the threshold to obtain a clear 
body boundary. 
Max. 
Function: set Max. as the 3D image rendering mode. 
Displays the maximum echo intensity in the observation 
direction. 
This is useful for viewing bony structures. 
Min. 
Function: set Min. as the 3D image rendering mode. 
Displays the minimum echo intensity in the observation 
direction. 
This is useful for viewing vessels and hollow structures. 
X-ray 
Function: set X-ray as the 3D image rendering mode. 
Displays the average value of all gray values in the ROI. 
X Ray: used for imaging tissues with different internal 
structures or tissues with tumors. 
iLive 
Function: iLive brings a better imaging experience by 
adding a light rendering effect to the traditional method. It 
supports the global lighting mode as well as the partial 
scattering mode, allowing human tissue texture to be 
revealed more clearly. 
5.13.3.3  Static 3D Image Viewing 
Enter/Exit Image Viewing 
  To enter image viewing: 
The system enters image viewing when image acquisition is complete. 
  Exit 
To return to 3D/4D image acquisition preparation status, press <Update> or <Freeze>. 
Activate MPR 
Click [VR/MPR] to activate MPR or 3D image (VR). 
MPR Viewing 
In the actual display, different colors for the window box and the section line are used to 
identify the MPR A, B and C.