MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
EN-25
Step
Display Line 2
Operation
Description
6.
Set the analog output value
using or , then press
to set the value and move to
D.Unt
Enter the weight value for which
100% analog output
has to be
generated.
Range: -1999 to 9999
7.
Set the display unit using
or , then press to set the
value and move to O.tYP
Value
Display unit
19
kg
20
g
21
t
22
Text of phys. unit/
preset via MP26/10
23
lb
8.
Set the analog output type
using or , then press
to set the value and move to
donE
Value
Output type
1
0 ... 20 mA continuous
2
4 ... 20 mA continuous
3
0...10V continuous
4
2...10V continuous
9.
Press to save it and move
to another menu level.
After pressing , the display
shows CAL. If you want to move
to the specific menu level for
further processing, press
when the display shows that
menu level. Otherwise, use the
up or down arrow key to move
to CALSCALEnd. Pressing
in the End menu level to exit
and return to normal weigh
mode/ display mode.