3 Operation Procedure
3
3 Operation Procedure
3.1 Power ON/OFF
This product is not equipped with a power switch.
TopowertheproductON/OFF,eitherswitchthepowersupplyitselfon/o󰀨,orconnect/disconnectthe
power cable.
When the power goes on, an ID number is auto
-
matically assigned, and then the product enters the
stand-by state.
Counting begins when the [SEL/CE] key is pressed
while the product is in the stand-by state.
Furthermore, when using multiple EJ Counters that are linked together, ID numbers are assigned in
sequence from the EJ Counter that is at the top when counter displays are viewed with the characters
oriented upright.
IDnumbersareassignedsequentially(gureshowsautomaticID
number assignments).
Tips
ID numbers from 01 to 08 are assigned automatically. If
arbitrary ID numbers (50 to 99) were set for Parameter
Number 19 when the product was last powered on,
those values are assigned instead.
If numeric display is selected for Parameter Number 09
when the product was last powered on, counting starts
immediately upon connection of INC model Linear
Gages.
When Linear Gages with origin mark are connected,
whether counting begins immediately at power-on or or
is deferred pending origin detection depends on the
setting made for Parameter Number 05.
For details, see
"3.3.2 List of Parameters" on page
32.
Counter stand-by state
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
Origin detection wait state
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1