3 Operation Procedure
3.2.5 Setting Preset/Tolerance Values
Preset and tolerance values can be set for each channel.
Tips
• Preset value setting is not possible if speed display is set as the display mode by parameter setting.
• Tolerance value setting is not possible if no tolerance is set by parameter setting. The tolerance value set-
tingsaredierentfor3-and5-steptolerancejudgment.Firstyoumustselect3-or5-steptolerancejudgment
in the parameter settings.
1
Display the channel for which you wish to set the preset and tolerance values.
Tips
For details, see "3.2.1 Switching the Display Channel (Ch.)" on page 22.
2
Press the [Fn] key.
» Display enters the preset value setting screen.
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1
3
Set the preset value.
Using the [MODE] (>) key to move between digits, change the digit values using the [P.SET] (^) key.
Thecurrentdigitashestoshowthatitisselected.
Ch.
TIR
Ch.
TIR
Ch.
TIR
Ch.
TIR
Ch.
TIR
Ch.
TIR
1
1
1
1
1
1
Movement between digits
[MODE]
(>) key
Changing numeric values
[P.SET]
(^) key
Setting a negative value
[P.SET]
(^) key
Tips
• While setting the preset values, is displayed by the tolerance indicator.
• If you wish to set a negative preset value, you can change the highest-order digit to a value with a minus
sign. The peak mode changes in the sequence shown below each time you press the [P.SET] (^) key.
0→1→2→……→9→-0→-1→-2→……→-9→0
• The direction in which digit selection and number setting proceeds can be reversed by pressing the [MODE]
(>) or [P.SET] (^) key together with the [Fn] key.
• You can omit steps
4
–
7
ifyouwishtonishaftersettingonlythepresetvalues.
4
Press the [SEL/CE] key.
» The preset value settings are applied.
» Display enters the tolerance value S1 setting screen.