32
No. 99MBC139A
3 Operation Procedure
3.3.2 List of Parameters
Tips
• Entry of incorrect parameter settings will prevent correct measurement results or result in display of an error.
• In case operation fails or an error is displayed, see "5 Troubleshooting" on page 51 to check
setting contents and make the correct settings.
■Basic parameters
Basic measurement parameters
Be sure to make these settings before starting measurement.
No. Set value
Per-
axis
setting
Set value: Operation
Default
value
Description
04
Linear Gage
resolution
(minimum
reading)
0: 0.005 mm (5 μm),
0.0002 in
1: 0.001 mm (1 μm),
0.00005 in
2: 0.0005 mm (0.5 μm),
0.00002 in
3: 0.0001 mm (0.1 μm),
0.000005 in
1
Set the minimum reading according
to the resolution of the connected
Linear Gage.
Make settings individually for the
A-axis and B-axis.
Tips
• Correct values will not be dis-
played if the settings do not
match the resolution of the
connected Linear Gage.
• The unit for minimum reading
is "in" when Parameter Number
22 is set to "1".
05
Origin detec-
tion function
N/A
0: Disabled
1: Enabled
0
Selects whether the origin function
is enabled or disabled when a Lin-
ear Gage with an origin point mark
is connected.
06
Counter
direction
0: + direction
1: - direction
0
Sets the relationship between the
direction in which the numeric value
changes and the direction of move-
ment of the Linear Gage spindle.
When 0 is set, the numeric value
increases as the spindle moves
toward top dead center.
When 1 is set, the numeric value
increases as the spindle moves
toward bottom dead center.
Make settings individually for the
A-axis and B-axis.
07
Origin detec-
tion direction
0: + direction
1: - direction
0
When a Linear Gage with an ori-
gin mark is connected, selects the
direction of the spindle of the Linear
Gage for origin detection.
Make settings individually for the
A-axis and B-axis.