25
No. 99MBC139A
3 Operation Procedure
3.2.3 Peak Hold
Inadditiontodisplayofthecurrentvalue,maximum/minimumvaluesandrun-outwidth(thedierence
between the maximum and minimum values) can be detected and displayed.
■Switching the peak mode
1
Display the channel for which you wish to switch the peak mode.
Tips
For details, see "3.2.1 Switching the Display Channel (Ch.)" on page 22.
2
Switch the peak mode.
The peak mode changes each time you press the [MODE] (>) key.
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
TIR MAXMIN TIR MAXMIN TIR MAXMIN TIR MAXMIN
MAXMIN
Current value display Maximum value display Minimum value display Run-out width display
■Clearing the peak values
1
Display the channel for which you wish to clear the peak values.
Tips
For details, see "3.2.1 Switching the Display Channel (Ch.)" on page 22.
2
Switch the peak mode to maximum, minimum or run-out value display.
The peak mode changes each time you press the
[MODE] (>) key.
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1
MAX
3
Press and hold the [Fn] key, and then press the [P.SET] (^) key.
» This clears the maximum and minimum values
and the run-out width.
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1
MAX