38
No. 99MBC139A
3 Operation Procedure
3.3.3 E󰀨ectofParameterChangesontheA-axisandB-axis
Changing the basic parameters initializes the settings and dependent channels of the associated axes.
Parameter changed
Setting initialized
Axis-related Channel-related
Number
display
Meaning Counter IC
Ori-
gin
Preset state, peak value
Preset value,
tolerance value
04 A Resolution (A-axis) (A-axis only) (channels dependent on A-axis only)
04 B Resolution (B-axis) (B-axis only) (channels dependent on B-axis only)
05
Origin function
selection
06 A
Counter direction
(A-axis)
(A-axis only)
(channels dependent on
A-axis only)
06 B
Counter direction
(B-axis)
(B-axis only)
(channels dependent on
B-axis only)
07 A
Origin detection
direction (A-axis)
(A-axis only)
(channels dependent on
A-axis only)
07 B
Origin detection
direction (B-axis)
(B-axis only)
(channels dependent on
B-axis only)
22 Unit
Tips
Channelsdisplayingsumanddi󰀨erence(A+BandA-B)aredependentonboththeA-axisandB-axis.
Since preset is not possible for the channel displaying speed, preset clear is not performed.