37
No. 99MBC139A
3 Operation Procedure
No. Set value
Per-
axis
setting
Set value: Operation
Default
value
Description
18
Hide the
lowest-order
digit.
N/A
0: Display all digits.
1: Hide the lowest-order
digit.
0
Hides the lowest-order digit. Howev-
er, data output to the optional inter-
face unit includes the lowest-order
digit.
19
Arbitrary ID
specication
N/A
00 to 49: ID numbers
assigned auto-
matically.
50 to 99: Arbitrary ID
numbers as-
signed (ID num-
bersspecied).
01
Set the ID number required when an
optional interface unit is connected.
When arbitrary ID numbers are set
in the range 50 to 99, the ID num-
bers set are assigned the next time
the power is turned on.
Tips
• When 00 to 49 is set, the ID num-
bers can be used until the power
isturnedo,butareautomatically
assigned in the range from 01 to 08
the next time the power is turned
on again.
• An error does not result if the same
arbitrary ID is set for multiple EJ
Counters that are linked together. If
ID numbers are duplicated during
use of the optional interface unit, a
communications error results.
20
Power saving
function
N/A
00: Display always lit
01 to 99: Display goes
out after the
speciedin-
terval passes
(specify the
interval length
in minutes).
00
Turningodisplayminimizespower
consumption.
Specify the time that elapses be-
tween when the last key is pressed
until the display goes out (01 to 99
minutes).
Whendisplayisturnedo,press
any key to turn it on again.
Tips
Counter operation for the axes, I/O
input and output, and data communi-
cation through the optional interface
unit continue even when the display
iso.
21
Parameter
initialization
N/A
0: Do not initialize.
1: Initialize.
0
When this parameter is set to 1, the
settings for all parameters other than
number 19 (the arbitrary ID) and
number 22 (the unit setting) are reset
to their default values. Further, the tol-
erance and preset values are cleared.
Tips
When you set 1 and press the [SEL/
CE] key, "cLEAr" is displayed and the
settings are initialized.