No. 99MBC139A
3 Operation Procedure
No. Set value
Per-
axis
setting
Set value: Operation
Default
value
Description
12
Origin re-de-
tection
(I/O input/out-
put setting)
N/A
0: Disabled
1:Onlyeectiveforthe
axis that is dependent
on the channel select-
ed by 1/2 SEL.
2: Axis dependent on
Ch. 1 and axis depen-
dent on Ch. 2.
0
When setting is enabled, origin
re-detection goes on stand-by when
the HOLD signal (I/O input) rises. If
the HOLD signal rises again while
original detection is on stand-by,
origin re-detection is canceled.
Tips
Thissettingaectsallaxesrelated
tothespeciedchannel.However,
this does not include the axis whose
channel is displaying speed.
Example:
• If the channel displaying the sum
(A+B)anddierence(A-B)isse-
lected, the A-axis and B-axis are
aected.
• If the channel displaying the B-axis
coordinates is selected, the B-axis is
aected
13
Preset by I/O
input
(I/O input
setting)
N/A
0: Executed only for the
channel selected by
1/2 SEL.
1: Executed for both
channels.
0
Selects whether preset processing
by external input is to be applied to
just one channel or both channels.
14
Chaect-
ed by the
CLEAR sig-
nal
(I/O input/out-
put setting)
N/A
0: Executed only for the
channel selected by
1/2 SEL.
1: Executed for both
channels.
0
Selects whether peak clear and
error clear are to be applied to just
thespeciedchannelortoboth
channels.
15
Peak value
preset
N/A
0: Disabled
1: Enabled
0
Sets whether the peak value is to
be set as the standard during peak
mode maximum value display or
minimum value display.
Tips
Peak value preset cannot be set
during run-out display.
16 Smoothing
N/A
0: No smoothing (update
at 5 ms intervals).
1: The average of 16
measurements is
displayed (update at
80 ms intervals).
2: The average of 32
measurements is
displayed (update at
160 ms intervals).
0
Counter values are averaged to
minimizeickerofthelowest-order
digit.
17
Speed sam-
pling cycle
N/A
0: 10 ms
1: 50 ms
2: 100 ms
0
Selects the sampling interval for
speed calculation.