26
No. 99MBC139A
3 Operation Procedure
3.2.4 Tolerance Judgment
Tolerance judgment is performed when it is enabled in the parameter settings.
There are two types of tolerance judgment: 3-step tolerance judgment and 5-step tolerance judgment.
Tips
Toperformtolerancejudgment,rstsetParameterNumber08to0(for3-stepparameterjudgment)or1(for
5-step parameter judgment), and then set the tolerance values.
For details, see "3.3 Parameter Setting" on page 30,. "3.2.5 Setting Preset/Tolerance Values" on page 27
■3-step tolerance
When S1 and S4 are set as tolerance values, 3-step tolerance judgment is performed as follows:
Judgment conditions Tolerance indicator
Measurement result < S1
(L1)
S1≤measurementresult≤S4
(L3)
S4 < measurement result
(L5)
Error resulting in no measurement
■5-step tolerance
When S1, S2, S3 and S4 are set as tolerance values, 5-step tolerance judgment is performed as fol-
lows:
Judgment conditions Tolerance indicator
Measurement result < S1
(L1)
S1≤measurementresult<S2
(L2)
S2≤measurementresult≤S3
(L3)
S3<measurementresult≤S4
(L4)
S4 < measurement result
(L5)
Error resulting in no measurement
■Target of tolerance judgment
Thetargetoftolerancejudgmentdi󰀨ersaccordingtopeakmode.
Peak mode Target of tolerance judgment
Current value display Current value
Maximum value display Maximum value
Minimum value display Minimum value
Run-out width display Run-out width