EasyManua.ls Logo

Mitutoyo EJ-102N - DC Power Supply Connection

Mitutoyo EJ-102N
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
No. 99MBC139A
2 Setup
2.3.2 DC Power Supply Connection
This section describes the procedure for DC power supply connection.
Tips
Prepare a power supply capable of supplying 10 V DC27 V DC (30 W).
Never use a power source that is shared with other equipment that draws high power.
WhenusingthisproductconnectedtoanLGF-seriesLinearGagewithresolutionof0.0001mm(0.1μm),please
usewithaDCpowernoiselter.(ThenoiselterusedfortestingbyMitutoyoistheCoselESP-06-472-D.)
For applicable Linear Gages, see "■LinearGageswith6-pinconnector"onpage4.
1
Connect the power cable to the provided connector plug.
Pin 1 (A): 10 V–27 V DC
Pin 2 (B): GND
C
A (10V–27V DC)B (GND)
D
●
Whenthecableisttedwithpinterminals
Push the pins all the way in to terminals A and B.
Tips
Compatible pin terminals are those with conductive ferrule sleeves that are 10 mm in length.
Recommended part: Phoenix Contact part number 3241129 AI 0,34-10 TQ
Use a power cable with a length of less than 30 m, and make a one-to-one connection between the EJ
Counter and an insulated power supply.
The EJ Counter internal ground is connected to ground internally.
●
When the cable ends in stranded wire
SlipthewiretipsallthewayintoterminalsAandBwhileusingatoolsuchasaat-headprecision
screwdriver to press connector plug buttons C and D, and then release the buttons.
Tug on the cable to make sure that the stranded wires are securely fastened to the connector plug.
●
Disconnecting the cable
Todisconnectthecablefromtheplug,pulloutwhilepressingthebuttonswithaat-headprecision
screwdriver or the like.
NOTICE
Forcibly pulling the connector plug out will result in damage to the connector or discontinuities in the cable.