41
No. 99MBC139A
4 Data Output
4
4 Data Output
4.1 I/O Function
By using the I/O connector, results of tolerance judgment can be output as signals, and settings of the
EJ Counter can be made by external control signals.
4.1.1 Connector Plug
Recommended connector plug:
y Phoenix Contact connector DFMC 1,5/ 5-ST-3,5-LR - 1790519 (with lock levers, standard acces-
sory)
y Phoenix Contact connector DFMC 1,5/ 5-STF-3,5 - 1790328 (with screw lock)
4.1.2 Wiring Procedure
1
Connect the cable to the connector plug.
A
B
●
Whenthecableisttedwithpinterminals
Push the pins all the way in to terminals A.
Tips
• Compatible pin terminals are those with conductive ferrule sleeves that are 10 mm in length.
Recommended part: Phoenix Contact part number 3241129 AI 0,34-10 TQ
• Use an input/output cable with a length of less than 30 m for I/O input/output.
●
When the cable ends in stranded wire
Slip the wire ends all the way into terminals B while pressing the connector plug buttons, and then re-
lease the buttons.
Tug on the cable to make sure that the stranded wires are securely fastened to the connector plug.