53
No. 99MBC139A
5 Troubleshooting
5.2.3 Error List
Display
Interface
unit
output
I/O
output
Cause Solution
Error 10 Yes Yes
Abnormal power
voltage
(Voltage is outside
of the range from
10 V through 27 V.)
Make connection to a power supply of suitable
voltage, and then do the error clear operation.
Error 12 Yes Yes
Internal memory
abnormality
(data abnormality)
Do the error clear operation.
Tips
If the error cannot be cleared, there may be an
internal malfunction. Please contact the nearest
Mitutoyo sales representative.
Error 13 Yes Yes
Internal memory
abnormality
(access abnormal-
ity)
Do the error clear operation. If the error cannot
becleared,cyclethepoweroandon.
Tips
If the error cannot be cleared, there may be an
internal malfunction. Please contact the nearest
Mitutoyo sales representative.
---------
Yes
(Error15)
Yes
Counter stand-by
state after pow-
er-on
Change to the counter display screen with the
[SEL/CE] key.
Tips
When Parameter Number 09 is set to 1 (counter
display), this error is not displayed.
Origin not acquired Do the error clear operation.
Power supply inter-
ruption
Verify power supply connection, and then do
the error clear operation.
Error 20 Yes Yes Excess speed
Verify measurement conditions (such as spin-
dle movement speed), and then do the error
clear operation.
Error 30 Yes Yes
Counter value
overow
After doing error clear operation, check posi-
tioning of the measurement work and spindle,
verify the preset values of the product, and
then set presets again.
Tips
• Presets are cleared automatically.
• The origin is not acquired.
Error 40 Yes Yes
Linear Gage dis-
connection de-
tected (other than
origin signal), or
excess speed
Verify connection of the Linear Gage. After
making proper connection, do the error clear
operation.
Verify measurement conditions (such as spin-
dle movement speed).
If the error cannot be cleared, there may be a
discontinuity in the cable.
Please contact the nearest Mitutoyo sales
representative.