54
No. 99MBC139A
5 Troubleshooting
Display
Interface
unit
output
I/O
output
Cause Solution
Error 41 Yes Yes No origin signal
When using a Linear Gage with origin mark,
check whether the origin signal is disconnect-
ed.
When using a Linear Gage without origin mark,
such as an INC model, set Parameter Number
05 to 0.
Tips
For details about Parameter Number 05, see
"3.3.2 List of Parameters" on page 32.
Error 60 Yes Yes
More than eight EJ
Counters are linked
together.
Turn the power back on after reducing the
number of linked EJ Counters to no more than
eight.
Tips
When multiple counters are linked together, errors
are displayed by the two end counters.
Error 80 Yes Yes
Peak detection
error
Verify measurement conditions (such as spin-
dle movement speed), and then do the error
clear operation.
Error 81 Yes Yes
Counter IC reset
error
Occurred because a Linear Gage was con-
nected or removed during operation.
Do the error clear operation.
Error 85 Yes Yes
Counter IC over-
ow
Electrical noise could be a problem.
Do the error clear operation.
Error 90 Yes Yes
Tolerance value
setting error
The tolerance value is not set correctly.
With 3-step tolerance judgment:
TheconditionS1≤S4isnotsatised.
With 5-step tolerance judgment:
TheconditionS1≤S2≤S3≤S4isnotsatis-
ed.
Press the [SEL/CE] key together with the [Fn]
key to return to preset/tolerance value setting
and reset the tolerance value.
Error 95 No No
A key was operated
with key protect
enabled.
Set Parameter Number 01 to 0 (no key protect).
Tips
For details about Parameter Number 01, see
"3.3.2 List of Parameters" on page 32.