EasyManua.ls Logo

Mitutoyo EJ-102N - Page 42

Mitutoyo EJ-102N
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
No. 99MBC139A
3 Operation Procedure
5
Set tolerance value S1 (tolerance indicator display: ).
Using the [MODE] (>) key to move between digits,
change the digit values using the [P.SET] (^) key.
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1
Tips
If you wish to set a negative tolerance value, you can change the highest-order digit to a minus sign.
The direction in which digit selection and number setting proceeds can be reversed by pressing the [MODE]
(>) or [P.SET] (^) key together with the [Fn] key.
6
Press the [SEL/CE] key.
» Tolerance value S1 will be applied.
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1
Setting 3-step tolerance
MODE
Fn
SEL
/
CE
Ch.
TIR
P.SET
1
Setting 5-step tolerance
» When setting 3-step tolerance values, display enters the tolerance value S4 (tolerance indicator display:
) setting screen.
» When setting 5-step tolerance values, display enters the tolerance value S2 (tolerance indicator display:
) setting screen.
Tips
The state of the tolerance indicator changes in the following sequence according to the tolerance value settings
(S1, S2, S3 and S4).
Tolerance value
Tolerance indicator
3-step tolerance value
setting
5-step tolerance value
setting
S1 S1
S2
S3
S4 S4
7
Repeat steps
5
to
6
to set tolerance value S4 (tolerance values S2, S3 and S4 when setting
5-step tolerance values).
Tips
Display returns to the preset value setting screen of step 2 when you press the [SEL/CE] key while in the toler-
ance value S4 setting screen.

Related product manuals